Bioquell’s new calibration capability reduces bio-decontamination cycle time by 20%

Published: 21-Nov-2019

More-accurate calibration of low-level hydrogen peroxide vapour instruments improves efficiencies and ensures employee safety

Bioquell has created the capability to calibrate low-level hydrogen peroxide vapour instruments accurately at one part per million (1ppm), helping to reduce cycle times by approximately 20%.

The company is now able to calibrate to a published Occupational Safety & Health Administration (OSHA) method at the critical concentration point, ensuring safety of employees while optimising the cycle.

It is important that any low-level hydrogen peroxide sensor, used to determine the end of a decontamination cycle, is accurate. This is to ensure the safety of the personnel entering the area, as well as to optimise the cycle time.

The present calibration methods used by sensor manufacturers are often not performed with hydrogen peroxide vapour, and are not conducted at a vapour concentration at the critical value of 1ppm.

Testing and experience have shown that inaccuracies can unnecessarily prolong decontamination cycles by up to 20%.

Bioquell now has a calibration capability able to expose these sensors to a 1ppm hydrogen peroxide vapour concentration verified accurately against the published OSHA 1019 method.

When decontaminating a room or enclosure, entry should be prohibited until the Occupational Exposure Limit (OEL), also known as Permissible Exposure Limit (PEL), has been reached so the area can then be safely released. The level is based on a time weighted average (TWA) over eight hours.

The removal of hydrogen peroxide from an enclosure is a logarithmic process, meaning that it takes a disproportionally long time to reduce from 2 to 1ppm compared to earlier in the cycle, therefore inaccuracy of 0.5ppm can have a significant effect on the cycle length.

To resolve this issue, Bioquell has created the capacity to calibrate accurately at 1ppm.

The calibration instrument produces a highly-accurate stream of hydrogen peroxide vapour at 1ppm concentration with better than ±0.1ppm accuracy.

However, this relates to the accuracy of the calibration system, unrelated to the accuracy or repeatability of the sensor itself.

Bioquell’s calibration method is highly automated to ensure accuracy, repeatability and cost-effectiveness, and checked against the OSHA method to ensuring ongoing accuracy.